Descargar Imprimir esta página

identités 822013 Manual De Instrucciones página 21

Publicidad

Idiomas disponibles

Idiomas disponibles

4. After assembly, insert the battery into the slot between the two front wheels.
Press any button on the panel to start the transfer chair. Before first use, please refer to the commissioning guide and
perform a reset.
Before using for the first time, please refer to the commissioning guide and perform a reset.
5. Install the accessories:
Patient protection belt; bed tray; bed tray support; seat cushion; battery.
PUTTING INTO SERVICE
1.
Before use, ensure that the battery is correctly installed and that the transfer chair is powered on.
2.
For the first time use, a reset operation is required. Before use, put the chair into standby mode in one of two ways:
a.
The chair will automatically go into sleep mode after 1 minute of inactivity.
b. Or press and hold the up and down buttons simultaneously for 3 seconds to force the chair to sleep mode.
3.
Reset: In standby mode (no indicator light is on), press and hold the up and down buttons simultaneously for 3 seconds
to force the reset. Please ensure that no one is moving during the force reset. While the force reset is in progress, if
there is a If there is an accident, press any button to stop the reset.
4.
During the hard reset, if an overload or voltage loss occurs, the circuit has a protective function, and the column lift will
stop immediately.
5.
When there is uncoordinated unilateral lifting, it is necessary to carry out inspections in unloaded conditions to find the
abnormal points.
6.
If an uncontrollable abnormality occurs, quickly remove the battery.
DMI-F-07_Annexe 2_Notice.2
21

Publicidad

loading