Washing with too much water, with jet pressure or in a washing tunnel.
Outdoor use or to transport a patient in a vehicle.
Displacement on soft unconditioned ground.
Movement with a patient on terrain with an inclination greater than 5º.
Extreme use of the motor without respecting the maximum load indicated on its label.
Use of accessories or equipment other than those recommended by the manufacturer.
Any other use of the crane not in accordance with its intended purpose.
Failure to respect the work cycle of the equipment: 2 minutes in continuous use and 18 minutes at rest.
5.- SLING (see images on page 6 and 7)
Tecnimoem recommends the harness supplied with the lift as it can be used by physically dependent patients,
although they must have a good degree of head control. It allows easy access to technical and auxiliary
personnel, and can be used both sitting and lying down.
100% Polyester Composition. The materials used to manufacture the harness are hypoallergenic . No skin
reactions have been described, however, if any, it is recommended to consult the prescriber.
Adjustment in three heights with safety anchor (N).
5.1.- Sling safety and care
The choice of the appropriate size as well as the correct placement of the product in the area to be treated are
essential to effectively fulfill its function, for its useful life and for its safe use.
SECURITY. Never use with loads higher than the maximum load of the crane (marked on it).
Check that the sling is in perfect condition before using it on patients. It is recommended to use it previously in
healthy people to familiarize yourself with it and check that it adapts to the size and characteristics of the
patient.
Do not use the harness if it shows signs of tearing, wear or deformation.
The suspension straps at the ends of the harness straps (blue color) must not be used. They are safety ties in
case of breakage of any of the other suspension straps (N).
CARE. Recommended washing at 30ºC with neutral soap. Do not bleach or dry clean. Does not support ironing.
Dry on flat surface. Do not expose to the sun or heat source. Does not support dryer.
5.2.- Placing the sling on a seated patient
Lean the patient forward, ensuring that they are well
supported to reduce the risk of falling and to convey
safety.
Slide the sling under the patient's back, in the center,
with the opening symmetrical to the axis of the spine.
Rest the patient's back against the sling (graphic 1). Pass the leg support to the front of the chair on each side
of the thigh and then pass each leg support under each thigh (chart 2)