Instructions for Use
Do not install or use the pump in an environment exposed to atmospheric agents
(rain, snow, ice), dust, aggressive gases, or in explosive environments or those with
a high fire risk.
If placing the IDP-3 pump inside an enclosure, provide ample room to supply
ambient air to both the front and rear air intakes of the pump.
During operation, the following environmental conditions must be maintained:
Temperature:
Relative humidity:
There are four versions of the IDP-3:
100 VAC, 50/60 Hz
115 VAC, 60 Hz
220-230 VAC, 50/60 Hz
24 VDC
WARNING
The pump must be installed in a way that allows an easy interruption of the
line voltage.
Be certain that your electrical mains power voltage corresponds to that indicated
CAUTION
on the production label which is on the rear of the pump.
For the:
AC version, connect the pump to the power supply using an IEC-320 style
power cord of at least 10 A capacity.
DC version, connect to the power supply using the power cord supplied with the
pump.
Never disturb the two hex head bolts on each side of the pump head
CAUTION
(see Figure 17). Disturbing these bolts will cause loss of performance and/or
pump damage.
144
+5 °C to +40 °C (41 °F to 104 °F)
0 to 95% (non-condensing)
Agilent IDP-3 Dry Scroll Vacuum Pump User Manual