Technical Information
Figure 21 Grounding Plug and Outlet
If the product must be reconnected for use on a different type of electric
WARNING
circuit, the connector should be replaced by qualified service personnel.
Grounding Instructions (DC version)
Ensure the 24 VDC power supply and 0-10 V power supply are grounded according
NOTE
to the correct practice.
Extension Cords
If you must use an extension cord with this product:
Agilent recommends using only extension cords with a minimum of 16-gage
wire and a maximum length of 25 feet (7.6 m).
Use only a 3-wire extension cord that will accept the plug.
Make sure your extension cord is in good condition.
Be sure the extension cord is rated high enough to carry the current your
product will draw. An undersized cord will cause a drop in line voltage resulting
in loss of power and overheating.
Agilent IDP-3 Dry Scroll Vacuum Pump User Manual
Grounded
Outlet
Box
Ground
Pin
Grounded
Outlet
165