Technical Information
Kits and Service Options
Agilent pumps will provide many years of trouble-free service if the maintenance
procedures and intervals are observed. Cleaning and tip seal replacement are
recommended when pump base pressure has risen to an unacceptably high level
for your application. If your pump exhibits humming or grinding noises from the
bearings, a major overhaul should be done by Agilent or an authorized rebuild
center. Advance exchange pumps are available to minimize downtime.
The parts needed for tip seal replacement on the IDP-3 are available in the kit
described in the following table. This kit contains seals and O-rings, and can be
obtained from your Agilent dealer.
Table 9 Tip Seal Replacement Kit
Part Number
IDP3TS
Tip Seal Replacement
The parts and tools required to replace tip seals are listed below:
Tip Seal Replacement Kit P/N IDP3TS
4 mm Allen wrench
Razor blade or side-cutting pliers
Compressed air (optional).
Use latex gloves, face mask and safety glasses when using compressed air.
NOTE
WARNING
If dangerous gases were being pumped, flush the pump with air or
inert gas for at least 10 minutes prior to disassembly.
174
Description
Replacement Tip Seal Set
Agilent IDP-3 Dry Scroll Vacuum Pump User Manual
Contents
Replacement Tip Seals and
O-rings for IDP-3 pumps.