Dräger PEX 3000 Manual Del Usuario página 30

Ocultar thumbs Ver también para PEX 3000:

Publicidad

Idiomas disponibles

Idiomas disponibles

en
|
Control
Measuring mode
Display of the measured value 0 to 99 or 0 to 9.9
Current output 4 to 20 mA
long
Key
▼M
(>3 s)
short (>1 s and <3 s)
ZE
Adjust
zero-point ▼
SP
Adjust
sensitivity
SL
Keys
Current output
Test 4-20 mA
▲+▼
4.0 mA
CL
Keys
Current output
Adjustment 4-20 mA
▲+▼
4.0 mA
CU
Keys
Current output
Set sensor current
▲+▼
3.4 mA
dP
Keys
Current output
Set decimal point
▲+▼
3.4 mA
So
Keys
Current output
Software Version
▲+▼
3.4 mA
AZ
Keys
Current output
Autozero
▲+▼
3.4 mA
If no key is pressed in any status condition for more than 4 minutes, the menu is closed automatically without saving
and the current measured value is displayed again.
30
Fault has
Fault
occurred
rectified
Fault
Display E0 to E8
Current output 1 mA
Keys
Current output
Calibrate
▲+▼
3.4 mA
zero-point
Keys
Current output
Calibrate
▲+▼
3.4 mA
sensitivity
Calibration menu
Set current
output to X mA
Calibrate
power signal
Calibrate
sensor current
Display decimal
point
Display the
software version
Autozero status
display
Maintenance menu
4.3
Calibration menu
To enter the calibration menu, press the ▼ key for more than one second but
less than three seconds. The display shows a flashing point on the bottom right
indicating the calibration mode.
4.3.1
ZE – Calibrate the zero-point
1. Navigate using the ▲ or ▼ keys until the display shows ZE ("ZEro") and
confirm by pressing the (▲+▼) keys. The output signal switches to 3.4 mA.
The display shows the currently measured gas concentration (negative
values are displayed up to "–5" or "–,9").
2. Apply zero gas to the sensor and wait for signal stabilisation (max.
3 minutes).
3. Press the ( +▼) keys to save the displayed value as a new zero-point.
4. Press the (▲+▼) keys once again to exit the function. The ZE display
appears once again.
5. Press the ▼ key to the calibrate the sensitivity or
6. Press the
button to switch back to the measuring mode.
4.3.2
SP – Calibrate the sensitivity
1. Navigate using the ▲ or ▼ keys until the display shows SP ("SPan") and
confirm by pressing the (▲+▼) keys.
|