Vacuson 40/60
3.
Rotate Turn and Tilting Lever by 180° (Turn and
Tilting Lever now facing away from the grasp).
Make sure the gripper catches the rim of the jar.
Turn and Tilting Lever in
locked position.
6.4
Device preparation
1. Prior to use all sterilizable parts (Tubes, Cannulas,
Cannula handlebar, char lid and bottles) must be
sterilized.
2. Attach secretion jars with mounted and locked lid
to the device.
3. Lay the pneumatic pedal on the floor and plug the
connection tube into the port for the pneumatic
pedal.
Operating Instructions, Vacuson 40/60, Ref. 31997, V43/13
4. Flap down Turn and Tilting Lever into the desig-
nated groove.
Turn and Tilting Lever locked
and secured.
4. The Vario-AIR-Pedal is connected to the air inlet
port at the control unit by the connection tube. If
not in use the port is covered with a lid. Make sure
to cover the port again when Vario-AIR-Pedal is
not connected.
5. Attach quiver (optional) at the mounts of the rolla-
ble stand (optional).
EN