RDO
Optical Dissolved Oxygen Technique
®
RDO Optical Dissolved Oxygen Temperature Display and
Calibration
RDO Optical Dissolved Oxygen Temperature Display
Star Plus meters allow the temperature to be viewed on individual measurement
lines in addition to the temperature display on the top, left of the screen .
To view the temperature for the RDO dissolved oxygen measurement line:
1 . In the measurement mode, press
arrow icon will point to the selected line .
2 . Press
/
be changed to display dissolved oxygen (% saturation), dissolved oxygen (mg/L),
barometric pressure (no icon), temperature (no icon) or a blank line .
RDO Optical Dissolved Oxygen Temperature Calibration
The temperature calibration mode of the Star Plus meter allows the temperature on
each measurement line to be manually adjusted .
To calibrate the temperature for the RDO dissolved oxygen measurement line:
1 . In the measurement mode, press
and press
2 . Press
to begin the calibration .
3 . When the reading stabilizes, the arrow icon and the first digit will flash . Enter
the temperature by pressing
the next digit .
4 . Press
to save and end the calibration .
EN-38
to change the value on the selected line . The bottom line can
/
until the temperature is displayed .
Thermo Scientific Orion Star
to choose the bottom display line . The
to choose the bottom measurement line
/
to adjust each digit and
and Star Plus Series Meter User Guide
™
to move to