ELECTROMAGNETIC COMPATIBILITY
Table 1 - For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration
electromagnetic emissions
The DEVICE is intended for use in the electromagnetic environment specified below.
The customer or the user of the DEVICE should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
N/A
CISPR 11
Harmonic emissions
N/A
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
N/A
IEC 61000-3-3
Electromagnetic environment – guidance
The DEVICE is suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
The DEVICE is suitable for use in all establishments
domestic included and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
34
Table 2 - For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration
electromagnetic immunity
The DEVICE is intended for use in the electromagnetic environment specified below. The customer or the user of the DEVICE
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
± 8 kV contact
Electrostatic discharge (ESD)
± 2 kV, ± 4 kV, ± 8 kV,
IEC 61000-4-2
± 15 kV air
Power frequency
(50/60 Hz)
3 A/m; 50 Hz or 60 Hz
magnetic field
IEC 61000-4-8
Electromagnetic
Compliance level
environment – guidance
Floors should be wood,
concrete or ceramic tile.
± 8 kV contact
If floors are covered with
± 2 kV, ± 4 kV, ± 8 kV,
synthetic material, the
± 15 kV air
relative humidity should be at
least 30 %.
Power frequency magnetic
fields should be at levels
characteristic of a typical
3 A/m; 50 Hz or 60 Hz
location in a typical
commercial or hospital
environment.
35