Technical details electrical components
NL
Controller box: C3con1-x, manufacturer Concens AS, Denmark
Actuator: Con 5, manufacturer Concens AS, Denmark
SV
Battery: C3 NiMH 24 V – 1400 mAh
Battery charger: C3-CHG-01-XX-02, Input voltage: 100–240VAC,
Class II, Input frequency: 50–60Hz, Maximum power rating:
DA
50VA, Output voltage: 24VDC, Output current: Max 400mA, IP22.
Expected service life of the battery is 5 years under heavy use
condition according to the manufacturer.
FI
IS
Guidance and manufacturer's declaration
Electromagnetic EMISSIONS for all ME EQUIPMENT and ME SYSTEMS
The TOPRO Taurus walking aid is intended for use in the electromagnetic environment specifi ed below.
FR
The customer or the user of the TOPRO Taurus walking aid should assure that it is used in such an environment.
Emissions test
IT
RF emissions
CISPR 11
ES
RF emissions
CS
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fl uctuations/
fl icker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration
Electromagnetic IMMUNITY for all ME EQUIPMENT and ME SYSTEMS
The TOPRO Taurus is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the TOPRO Taurus should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic fi eld
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Art. No. User Manual: 104801 – Revision C, 2022-05
Compliance
Group 1
Class B
Complies
Complies
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
Duty Cycle: 10% maximum activation (on) time 2 minutes
and the minimum deactivation (off ) time 18 minutes.
Working ambient temperature for actuator: −5°C to +40°C
Working ambient temperature for controller and battery
charger: +5°C to +40°C
Humidity: 20–70 %
Delivery and storing ambient temperature for actuator,
controller and battery charger: −20°C to +50°C
Humidity: 20–70 %
Electromagnetic environment – guidance
The TOPRO Taurus walking aid uses RF energy only for
its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
The TOPRO Taurus walking aid is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low voltage power
supply network that supplies buildings used for domestic
purposes.
Compliance level
Electromagnetic environment – guidance
± 6 kV contact
Floors should be wood, concrete or ceramic
± 8 kV air
tile. If fl oors are covered with synthetic
material, the relative humidity should be at
least 30 %.
± 2 kV for power supply
Main power quality should be that of a typical
lines
commercial or hospital environment.
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
Main power quality should be that of a typical
± 2 kV line(s) to earth
commercial or hospital environment.
<5 % UT
Main power quality should be that of a typical
(>95 % dip in UT)
commercial or hospital environment.
for 0,5 cycle
If the user of the TOPRO Taurus requires
40 % UT
continued operation during power mains
(60 % dip in UT)
interruptions, it is recommended that the
for 5 cycles
TOPRO Taurus be powered from an uninterrupti-
70 % UT
ble power supply or a battery.
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
Power frequency magnetic fi elds should be
at levels characteristic of a typical location
in a typical commercial or hospital environment.
–
56
–
TOPRO Taurus E