MEGABRAS MPK102e Manual De Uso página 12

Micro-ohmímetro digital

Publicidad

Idiomas disponibles

Idiomas disponibles

3.5. Display messages
When turning the equipment on using the
On/Off switch, this introduction message
appears for a while.
This message appears each time the
equipment needs to adjust any parameter
in order to optimize the readings.
Warns the operator that to be able to start
a test, the Test current control must be
at the zero position (fully counter-
clockwise). Otherwise, Start button will be
inhibited.
The equipment is ready to start a
measurement, thus the operator has to
press the Start button.
Indicates that the test current is not
enough to carry out the reading. It
appears at the beginning of each test and
it keeps on being there up to the operator
rotates the Test current control
clockwise.
Indicates that the measured resistance is
higher than the maximum value readable
in the selected range.
Indicates that some part of the equipment
has achieved the critical temperature.
Thus the system will cut the high current
generation.