SPECIFICATIONS
Info: Specifications are subject to change without prior notice.
1
Display Type
OLED display
2
SpO2
Display range: 0 - 100%
Measurement range: 70 - 100%
Accuracy: 70%-100%: ±2%;
0% ~ 69% no definition
Resolution: 1%
3
Pulse Rate
Display range: 0BPM ~ 250BPM
Measure range: 30BPM - 250BPM
Accuracy: 30 ~ 99BPM, ±2BPM;
100BPM ~ 250BPM, ±2%
Resolution: 1BPM
4
Probe LED Specifications
Wavelength
RED
660 ±2nm
IR
905±10nm
Info: Wavelength range information can be especially
useful to clinicians.
5
Power Requirements
Two AAA alkaline Batteries
Power consumption: Less than 30mA
Battery Life: Two AAA 1.5V, 600mA alkaline batteries could
be continuously operated as long as 30 hours
10
cs@grahamfield.com
Radiant Power
3.2mW
2.4mW
6
Environmental Requirements
Operation Temperature: 41°F ~ 104°F (5°C ~ 40°C)
Storage Temperature: -13°F ~ 158°F (-25°C ~ 70°C)
Ambient Humidity:
≤80% no condensation in operation
≤93% no condensation in storage
Atmosphere pressure: 86kPa ~ 106kPa
7
Classification
According to the type of protection against electric shock:
INTERNALLY POWERED EQUIPMENT;
According to the degree of protection against electric
shock: TYPE BF APPLIED PART;
According to the degree of protection against ingress of
water: IP22
According to the mode of operation:
CONTINUOUS OPERATION