TROUBLESHOOTING
Problem
Possible Problems
SpO2 or PR cannot
1. Finger is not inserted correctly
be shown normally
2. Patient's SpO2 value is too low
SpO2 or PR is
1. Finger might not be inserted deep enough
shown unstably
2. Excessive patient movement
Oximeter cannot
1. No batteries or low battery power
be powered on
2. Batteries might be installed incorrectly
3. The oximeter might be damaged
Indication lamps
1. The product automatically powers off when no
are suddenly off
2. Battery power is too low
"Error3" or
1. Err 3 means the red emission LED is damaged
"Error4" displays
2. Err 4 means the infra-red emission LED is
"Error6" displays
Err 6 means the screen has failed
"Error7" displays
Err 7 means the emission LED or reception diode is
damaged
AFTER SALES SUPPORT
800-347-5678
to be measured
signal is detected for more than 8 seconds
damaged
Solution
1. Retry by re-inserting the finger
2. There is excessive illumination
3. Try again. If you can verify there is no problem
with the product, go to a hospital immediately
for exact diagnosis.
1. Retry by re-inserting the finger
2. Remain still and calm
1. Replace batteries
2. Reinstall batteries
3. Contact Customer Service at 800-347-5678
1. Normal
2. Replace batteries
1. Check red emission LED
2. Check infra-red emission LED
Contact Customer Service at 800-347-5678
Contact Customer Service at 800-347-5678
9