GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY - FOR ALL EQUIPMENT AND SYSTEMS
THAT ARE NOT LIFE-SUPPORTING
The 5 3 4 9 7 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the
user of the 53497 Pulse Oximeter should assure that it is used in such an environment.
Immunity
IEC
60601
Test
Test Level
Radiated
3 V/m
RF
80 MHz to
IEC
2.5 GHz
61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations, Electromagnetic propagation is affected by absorption and reflection
structures, objects and people.
AFTER SALES SUPPORT
800-347-5678
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Compliance
Electromagnetic Environment – guidance
Level
Portable and mobile RF communications equipment should be used no closer
3 V/m
to any part of the Pulse Oximeter (53497), including cables, than the
recommended separation distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
frequency range
Interference may occur in the vicinity of equipment marked with following
symbol:
80 MHz to 800 MHz
800 MHz to 2.5 GHz
, should be less than the compliance level in each
a
.
b