Info: After turning on the Oximeter, each time you press
the power switch, the Oximeter will switch to another
display mode. There are 6 display modes shown as follows:
8
After no or low signal has been received for approximately
eight seconds, the pulse oximeter will automatically shut
off.
FRONT PANEL
Oxygen Saturation
Pulse Rate
Pulse Oximeter
SpO
% Waveform
Power Switch
2
The bar graph height indicates the pulse and signal strength.
The bar should be greater than 30% for a proper reading.
AFTER SALES SUPPORT
800-347-5678
Power Low Indicator
Pulse Oximeter
Pulse Bar Graph
INACCURATE MEASUREMENTS MAY BE CAUSED BY
1
Significant levels of dysfunctional hemoglobin (such as
carbonyl - hemoglobin or methemoglobin).
2
Intravascular dyes such as indocyanine green or methylene
blue.
3
High ambient light. Shield the sensor area if necessary.
4
Excessive patient movement.
5
High-frequency electrosurgical interference and
defibrillators.
6
Venous pulsations.
7
Placement of a sensor on an extremity with a blood
pressure cuff, arterial catheter, or intravascular line.
8
The patient has hypotension, severe vasoconstriction,
severe anemia, or hypothermia.
9
The patient is in cardiac arrest or is in shock.
10 Fingernail polish or false fingernails.
11 Weak pulse quality (low perfusion).
12 Low hemoglobin.