MEASUREMENT WITH SET PRESSURE
The UA-1030T is designed to detect the pulse and to inflate the cuff to a systolic
pressure level automatically. If re-inflation occurs repeatedly, use the following
methods.
During the blood pressure measurement, re-inflation may occur. A fixed pressure
value can be set to avoid re-inflation.
1. At step 6 of the clock adjustment procedure (see page
E-10), press the
button to go to the pressure setting
SET
mode. The current setting blinks.
2. Press the
or
button to select a pressure value
about 30 mmHg or more above your expected systolic
pressure from the following.
: Automatic pressurization (default value)
AUTO
: Pressure value of 180 mmHg (fixed)
180
: Pressure value of 210 mmHg (fixed)
210
: Pressure value of 240 mmHg (fixed)
240
3. Press the
button to go to the TriCheck setting mode.
SET
Press the
button to turn the device off. After three minutes of
START
non-operation, the device will turn off automatically. The next measurement
will be performed with the new pressure value.
MEASUREMENT WITH DESIRED SYSTOLIC PRESSURE
Use this method when re-inflation occurs repeatedly even if the pressure value is
set to
in the procedure above or when the results are not displayed even if
240
the pressure decreases to 20 mmHg or less.
1. Place the cuff on the arm (preferably the left arm).
2. Press and hold the
button until a number
START
about 30 to 40 mmHg higher than your expected
systolic pressure appears.
3. When the desired number is reached, release the
button to start measurement. Continue to
START
measure your blood pressure as described on the
previous page.
E-15
or
or
TriCheck
Turn off
setting mode
TriCheck
Turn off
setting mode
At heart level
Press and hold the button
START
Release the button
At heart level
at the desired
systolic pressure
Press and hold the button
START
Release the button
Refer to previous page
at the desired
systolic pressure
for measurement
Refer to previous page
for measurement
A&D Medical (Toll-Free): 1-888-726-9966
1932-1AIc UA787EXC•06 illust•03,12p symbols.FH7
123
The UA-1030T blood pressure monitor provides a blood pressure and pulse rate
123
measurement even when an irregular heartbeat occurs. The Irregular Heartbeat
123
symbol
will appear in the display window in the event an Irregular
123
123
Heartbeat has occured during measurement. An irregular heartbeat is defined as
err CUf
123
a heartbeat that varies by 25% from the average of all heartbeat intervals during
123
the blood pressure measurement. It is important that you relax, remain still and
123
refrain from talking during measurements.
123
123
Note: We recommend contacting your physician if you see this symbol frequently.
123
123
The Cuff Inflation Meter is located on the left side of the display screen to tell
you when the blood pressure monitor is inflating and deflating the cuff. The Cuff
Inflation Meter moves up during inflation and moves down during deflation.
Inflation in Progress
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
WHAT IS AN IRREGULAR HEARTBEAT
p-3
ABOUT CUFF INFLATION METER
Deflation/Measurement
Inflation Complete
in Progress
E-16