Patient target group
The patients themselves are not in direct contact with the product. The patient target
group is therefore defined by the device connected.
When using the AGSS (anaesthetic gas scavenging system) Forano as intended, the
focus is not on protecting the patient but rather on protecting persons in the immediate
vicinity. The medical personnel in particular are affected.
Intended users
The medical device is intended for clinical use by medically qualified personnel
(therapists, care personnel, doctors).
Operation
Before starting operation each time, an external visual inspection and a functional test
have to be carried out. After visual inspection for obvious damage or contamination, the
actuator must be activated and smooth movement verified.
To establish a connection, an AGSS plug or a hose with an AGSS type 1L connection is
inserted into the terminal unit.
By pressing the actuator on the terminal unit, the connector is released from the terminal
unit. The plug should be held by hand! The suction is interrupted.
Cleaning
The terminal unit can be wet cleaned with commercially available disinfectants and
cleaning agents. When cleaning, make sure that no liquid enters the terminal unit. The
terminal unit must be completely dry prior to subsequent use. The hygiene plan of the
healthcare facility must be observed.
Installation
•
The installation housing in the hollow-wall / under plaster must first be secured in
the wall!
•
Hollow-wall installation: Fasten the housing with six screws left and right, solder the
pipes to the piping system
•
Under plaster: Plaster the housing in the wall, solder the pipes to the piping system
•
Connect the terminal unit to the pipe or ejector connections using the connecting
lines (ejector can be plugged and fixed with a screw!)
•
Push the terminal unit with the front panel into the housing and fasten it with two
screws
•
Supply systems: Fix the terminal unit with two screws, insert the ejector into the
exhaust air line and secure it with a nut
•
Connect the terminal unit, ejector and pipe connections to the connection lines
•
Adjust suction capacity with valve spindle (11 o'clock position) with hexagon fittings
•
(2,5 mm)
•
Press the actuator on the terminal unit
•
Check anesthesia gas exhaust system according to DIN EN ISO 7396-2