Maintenance and inspection
Always perform a visual inspection for external damage before using the terminal units
(external damage). A function and tightness check should be performed at least once a
year (if necessary: replacement of wearing parts), maintenance should be carried out every
six years (measurement and safety check) in particularly high-use areas more frequently!
Faults and damage must only be dealt with by authorised expert personnel. Only use
genuine GREGGERSEN replacement parts when performing repairs.
Disposal
The terminal unit is to be disposed of in accordance with the relevant national regulations.
Technical Data
Design:
Connection:
Throughput:
Gas consumption:
Installation:
Our products are partly made of brass with a lead content of more than
0.1%. As an alloy component, lead is firmly bound in the material so that no
exposure is to be expected.
13
All-metal design, pluggable ejector for vacuum generation,
single-handed operation for connecting/disconnecting,
indicator for operational control, integrated valve for
throughput adjustment.
Supply lines underplaster/ hollow wall/ onplaster
Pressure gas line:
Exhaust line:
Terminal unit / ejector
Compressed gas supply:
Ejector:
Anaesthetic gas dissipation: 15 mm spiral hose
Indicator:
min. 50 l/min at 500 kPa compressed air
approx. 12 l/min at suction throughput of 50 l/min
Concealed / hollow-wall installation
Ceiling supply systems
Wall-mounted supply systems
Cu pipe 8x1
Cu pipe 15x1
Connection for 4 mm PA hose
or 7 mm barbed hose fitting
Connection for 4 mm PA hose
or 7 mm barbed hose fitting
Connection for 4 mm PA hose
www.greggersen.com