BEFORE TESTING
Calibration for Testing
You must calibrate the meter every time you begin to use a new box of
total cholesterol test strips by setting the meter with the correct code.
Test results may be inaccurate if the code number displayed on the
monitor does not match the number printed on the individual foil
packet.
How to Code Your Meter
1. Insert the code strip when the
monitor is off. Wait until the code
number appears on the display.
NOTICE:
Make sure the code number on display,
code strip, and the individual foil package
are the same. The code strip should be
within the expiration date; otherwise, an
error message may appear.
2. Remove the code strip, the display
will show "OFF". This tells you that the
meter has finished coding and is ready
for a test.
13