Descargar Imprimir esta página

Fora Test N'Go Advance Voice Manual Del Propietário página 41

Publicidad

Idiomas disponibles

Idiomas disponibles

Error Messages
MESSAGE
Appears when the batteries
E-b
cannot provide enough power
for a test.
E-2
Expired code strip.
Appears when a used test
E-U
strip is inserted.
E-E
Problem in operation.
E-0
E-A
Appears when test strip is
E-F
removed while counting down.
Appears when the wrong code
strip is inserted or other
E-C
coding errors.
Appears when ambient
E-t
temperature is below or above
system operation range.
WHAT IT MEANS
WHAT TO DO
Replace the batteries
immediately.
Repeat the test with a new lot
of test strip.
Repeat with a new test strip.
Repeat the test with a new test
strip. If the meter still does not
work, please contact the
customer service for
assistance.
Repeat the test with a new test
strip.
Make sure to insert the right
code strip for the meter.
Check if the code numbers on
the code strip and the strip vial
label or individual foil package
are the same.
System operation range is 50°F
to 104°F (10°C to 40°C), 10%
to 85%. Repeat the test after
the meter and test strip are in
the above temperature range.
33

Publicidad

loading

Este manual también es adecuado para:

Td-4183fTest n'go advance