the absorbent hole of the test strip touches the drop. Once the
confirmation window fills completely, the meter will begin counting
down.
NOTE:
For the control solution test, your device will tag this measurement as QC test
automatically.
To avoid contaminating the control solution, do not directly apply the control
solution onto the strip.
3. Read and compare the results
After the meter counts to 0, the result of the control solution test will
appear. Compare the result with the range printed on the test strip vial
or individual foil package. The result should fall within this range. If
not, please read the instructions again and repeat the control solution
test.
With "QC" displayed, the meter will store your test result in memory
under "QC".
Out-of-range results
If you continue to have test results fall outside the range printed on
the test strip vial or individual foil package, the meter and strips may
not be working properly. Do NOT test your blood, and call customer
service for help.
16