11.
Replacement fuse
Precautions
• Perform the procedures below with the equipment turned off.
• Disconnect the equipment from the mains supply and remove
the power cord.
• Disconnect the test leads.
• Disconnect the equipment from the mains, unplugging the power cord.
• With a screwdriver, remove the fuse holder cover, lightly pressing the lid
and turning ¼ counter clockwise.
• Remove the blown fuse and insert a new fuse in the fuse holder cover.
• Place the cover in the compartment and with the screwdriver turn ¼
clockwise or until the fuse holder cover properly fits the fuse
compartment.
Fuse Schurter, model SPT 5x20 (Time-lag) 15A/250V. High breaking capacity.
To replace the fuse, use only the fuse type specified in the
equipment.
12. Cleaning
Cleaning of this instrument should be carried out using a soft cleaning
liquid, after verifying that it doesn't affect the plastic parts used in the case
and in the Control Panel of this equipment.
To avoid electrical shock, make sure that the equipment is
completely dry before Power it On.
22