Descargar Imprimir esta página

MEGABRAS TR8703 Manual De Uso página 23

Medidor de relación de transformación

Publicidad

Idiomas disponibles

Idiomas disponibles

CONT/Ratio: Performs a Continuity test followed by Ratio test only in
VT/PT. Continuity test is not available (N/A) in CT mode.
Ratio test: Performs only the Ratio test in the VT/PT or CT mode.
NOTE: Test Mode can also be selected from the Main Screen by
pressing ENTER, which toggles the selection of CONT and Ratio.
3.3.5. Select storage mode
1.
From the top level of Configure Instrument, press the ▼ or ▲ key
until Select Storage Mode appears on the display, then press
ENTER.
2.
Select Auto or Manual using ▼ or ▲ and press ENTER. The display
will return to Configure Instrument > Select Storage Mode.
NOTE:
• In Manual mode, the user has a choice at the end of a test whether to
store the Measurement Record or not. The user also has a choice of
where to store the Measurement Record. So it is possible for the user
to store the Measurement Record at various locations leaving gaps in
the storage memory map.
All Measurement Records will be downloaded no matter where they
are stored in the memory.
• In Auto mode, the instrument will store the Measurement Record at the
end of a test starting with the first available location. All the subsequent
Measurement Records will be stored sequentially.
If the previously stored Measurement Record is encountered, the Auto
store mode will skip over and store the new Measurement Record in
the next available location.
24

Publicidad

loading