1. When putting on the cuff, spread the cuff and wrap it
around the upper arm evenly to appropriate tightness.
The correct cuff position refers to figure 2.5.
2. Connect the hose from the cuff to the connector
on the upper-side of the device with
the mark of "NIBP".
3. Pressing start/cancel button
pressure measurement.
Safety Instructions for blood pressure measurement
 Blood pressure measurement is prohibited to those who have severe hemorrhagic tendency
or with sickle cell disease, for partial bleeding may cause.
 Appropriate cuff should be selected according to the age and arm circumference
of the subject. Its width should be 2/3 of the length of the upper arm. The inflatable part
should be long enough to permit wrapping appropriately 80% of the limb. See the table
below for the dimensions:
Cuff Model
Child cuff
Small-sized Adult Cuff
Middle-sized Adult Cuff
Large-sized Adult Cuff
to start blood
Arm Circumference
17cm~22cm
22cm~30cm
30cm~42cm
42cm~48cm
ENGLISH
Figure 2.5 Cuff position