4.4 Preset programs
Programs (Cycles)
Program
1
Program type
Unpacked
Sterilization temp
1134ºC
Sterilization pressure
204kPa
Sterilization
4 min
Drying
9 min
Vacuum pulses
1 vacuum
Program
Program type
Sterilization temp
Sterilization pressure
Sterilization
Drying
Vacuum pulses
5. Control Panel
2
3
5.1 Controlling Panel
5.1.1 Integrated display
A. Display the chamber pressure, unit is: bar
B. Display the chamber temperature, unit is: ºC
C. Display the cycle situations. Please see the circumstance which is attached (Sterilizer operation cycle display chart).
D. Display the alarm error code when the sterilization gives the alarm during the working cycle, then you can check and find the
malfunction components according to the error code.
5.1.2 Increase value key (2)
UP choosing has the functions that are used for setting down the cycle program and filling in water pump, Press this button you
can choose the following programs, and if press button for 3 seconds, the machine would enter into filling in water automatically
program, and the LCD would display the working cycle state.
5.1.3 Decrease value key (3)
DOWN choosing has the cycle program, press this button you can also choose the following programs.
J
.P. SELECTA s.a.u.
CODIGO 80416 REV D
2
Packed
134ºC
204kPa
5 min
9 min
3 vacuums
6
Liquids
121ºC
104kPa
30 min
1 min
0 vacuums
1 Display
2 Increase value key
1
3 Decrease value key
4 Program key
5 Start/Stop key
4
5
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
3
4
Prion
Rubber
134ºC
121ºC
204kPa
104kPa
18 min
20 min
9 min
9 min
3 vacuums
1 vacuums
7
8
User
Test B&D
134ºC
134ºC
204kPa
204kPa
4 min
3,5 min
3 min
3 min
3 vacuums
3 vacuums
12/2023
5
Textile
121ºC
104kPa
20 min
18 min
3 vacuums
9
Test vacuum
134ºC
204kPa
5 min - 5 min - 10 min
Tel 34 937 700 877 Fax 34 937 702 362