Descargar Imprimir esta página

J.P. SELECTA AUTESTER ST DRY PV-B Manual De Instrucciones página 41

Autoclave clase b para esterilización 18l/ 23l con impresora integrada

Publicidad

Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES
5.3 Window of Sterilization Process
Some examples of PACKED program screens:
Program example with 3 pre-vacuum: PACKED
3 times pre-vacuum program example: PACKED
PACKED 134
3
-0.02bar
Preheating
PACKED 134
3
1.13bar
First pressurizing
PACKED 134
3
1.12bar
Second pressurizing
PACKED 134
3
1.74bar
Third pressurizing
PACKED 134
3
3:27
0.73bar
Exhausting and Drying
No. ASS0010
J
.P. SELECTA s.a.u.
CODIGO 80416 REV D
℃/2.1ba
1:04
∑1:20
70.6℃
℃/2.1ba
2:18
∑13:07
121.6℃
℃/2.1ba
2:18
∑18:09
121.6℃
℃/2.1ba
2:32
∑23:13
130.6℃
℃/2.1ba
∑26:13
108.6℃
REV-D
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
Steam Sterilizer Operation Manual
PACKED 134
3
3:24
-0.84bar
First vacuum
PACKED 134
3
3:18
-0.84bar
Second vacuum
PACKED 134
3
4:02
-0.82bar
Third vacuum
PACKED 134
3
3:27
2.13bar
Sterilizing
END
(00016)
053:12
-0.01bar
END
Tel 34 937 700 877 Fax 34 937 702 362
12/2023
Pag.: 41
℃/2.1ba
∑8:04
80.6℃
℃/2.1ba
∑16:04
80.6℃
℃/2.1ba
∑20:13
80.6℃
℃/2.1ba
∑26:13
134.6℃
35:30
90.6℃
Page 26 of 48