Quantum of leakage
15
Leakage forbidden
16
Protective earthing
impedance
17
Successional current
leakage under
the work temperature
18
Dielectric strength
with
working temperature
19
Empty load
20
Hollow load
21
Dryness, solid and wrapped
load
J
.P. SELECTA s.a.u.
CODIGO 80416 REV D
In the condition of the vacuum - 0.07 Mpa, the sterilizer shouldn't leak
0.0013Mpa within ten min.
The sterilizer can't leak under the work pressure
The impedance between protective earthing point of the power input faucet
and protective earthing can be touched all metallic parts, doesn't over than
0.1Ω.
a) Earth leakage current on the normal condition: ≤0.5mA the single blooey
state: ≤1 mA
b) Crust leakage current on the normal condition: ≤0.1 mA the single blooey
state: ≤0.5mA
a) A-al: It should bear the sine wave test alternative voltage ,50Hz, 1500v,
which between the web power input port and protective earthing can be
touched all metallic parts. It lasts 1 min, and hasn't the phenomenon of
breakage and flashover
b) A-a2: It should bear the sine wave test alternative voltage, 50Hz, 1500v,
which between the web power input port and the enclosure of which isn't be
pretended earthing. It lasts 1 min, and hasn't the phenomenon of breakage
and flashover
For all loads except hollow load A, the presence of saturated steam in the
usable space and the load is deemed to have been achieved when, through-
out the holding time, all temperatures measured in the usable space and the
load: (Attention: the theory of steam temperature is accounted by measur-
ing pressure, which can be considered the test temperature.)
are not lower than the sterilization temperature;
are not more than 4 K above the sterilization temperature;
do not differ from each other by more than 2 K.
The usable place temperature during the no-load can not over than the
scope of highest temperature.
For hollow load A and B, in order to confirm the presence or absence of
saturated steam, discriminate whether the indication system change in ac-
cordance with the system manufacturer predetermined color.
For wrapped load, any remaining moisture should not lead to wet packages
and shall not result in detrimental effects on the sterilizer load. Any remain-
ing water droplets on the inner side of pouch should evaporate within 5 min.
For solid load the moisture content should not exceed 0,2 %