5. Commissioning and operation / 6. Faults
1. Unscrew the transparent protective cap of the particle filter.
2. Screw off the particle filter.
3. Clean the particle filter with compressed air.
If a cleaning of the filter is not possible (no compressed air pistol available or too
heavily contaminated), screw out the filter and replace it.
CAUTION!
Damage to the sensor through incorrect cleaning
Overpressure in the console case can damage the sensor.
4. Tighten the protective cap again.
6. Faults
CAUTION!
Physical injuries and damage to property and the environment
If faults cannot be eliminated by means of the listed measures, the
instrument must be taken out of operation immediately.
Ensure that pressure or signal is no longer present and protect against
accidental commissioning.
Contact the manufacturer.
If a return is needed, please follow the instructions given in chapter
8.2 "Return".
For contact details, see chapter 1 "General information" or the back page
of the operating instructions.
Faults
Gas presence
detector indicates
gas concentration in a
gas-free area
Undefined reaction of the
operator screen
Weak suction capacity
of the gas presence
detector
WIKA operating instructions gas presence detector, model GIR-10
No pressure should be applied at the connection hose.
Causes
Zero point setting no longer
valid
Installed firmware is faulty
Blocked particle filter
Measures
Carry out a zero point setting, see
"
chapter 5.5.1
Zero point setting
Reset gas presence detector, see
"
chapter 5.5.2
Resetting
Clean or replace particle filter, see
"
chapter 5.5.4
Cleaning or replacing
"
the particle filter