Descargar Imprimir esta página

WIKA GIR-10 Manual De Instrucciones página 20

Ocultar thumbs Ver también para GIR-10:

Publicidad

Idiomas disponibles

Idiomas disponibles

7. Maintenance, cleaning and calibration
7. Maintenance, cleaning and calibration
7.1 Maintenance
This gas presence detector is maintenance-free.
EN
Repairs must only be carried out by the manufacturer.
This does not apply to the rechargeable battery replacement.
7.2 Cleaning
Suitable cleaning agents
Water
Conventional dishwashing detergent
Cleaning the instrument
Prior to cleaning, remove the power supply unit.
1.
Wipe the instrument surface using a soft, damp cloth.
2.
Electrical connections must not come into contact with moisture!
7.3 Calibration
We recommend that the instrument is regularly calibrated by the manufacturer with time
intervals of approx. 1,200 hours of operation or at the latest after 2 years. The default
settings can be corrected if necessary.
In addition, the gas presence detector can be calibrated by the operator in order to
reach a maximum accuracy in the critical range.
20
CAUTION!
Physical injuries and damage to property and the environment
Improper cleaning may lead to physical injuries and damage to property
and the environment. Decomposition products in the dismounted
instrument can result in a risk to persons, the environment and equipment.
Carry out the cleaning process as described below.
CAUTION!
Unsuitable cleaning agents
Cleaning with unsuitable cleaning agents may damage the
instrument and the product label.
Do not use any aggressive cleaning agents.
Do not use any hard or pointed objects.
Do not use any abrasive cloths or sponges.
WIKA operating instructions gas presence detector, model GIR-10

Publicidad

loading