Getting started
There are two ways of working with the IDA-6:
Run a procedure. A procedure is a set of measurements and tasks
that are defined in advance, using OneQA on a PC.
Visit OneQA.com to learn more.
Perform a measurement session. Select between three
measurement types: Flow & volume, Occlusion, or Bolus & PCA.
Add set values and limits to get a pass or fail result.
Measurement setup
With IDA-6 you can test up to four infusion devices at once.
1. Connect the included power adapter and turn on the product.
2. Connect a drain tube to the outlet port of a channel. Lead the other
end of the drain tube to a sink or leave it in a container on the table
to collect the fluid.
3. Connect an infusion device to the inlet port on the front side of the
product. Use distilled or deionized water.
4. Let the infusion device fill the administration set and the channel with
fluid. The internal volume of the channel is about 1 mL.