S19-921
Installation
Step 1: Mount Eyewash Pedestal to Wall (optional)
CAUTION: Eyewash station weighs 60 lb when filled to capacity.
Due to this weight, assistance from another
individual is required to safely install the unit.
Mounting bracket screws and wall anchors must be
strong enough to support the filled unit.
1 . To mount the pedestal to the wall, follow the procedure
outlined below:
• Secure the mounting bracket to the wall with screws
and wall anchors (provided by installer) .
2 . Attach the pedestal to the mounting bracket at the
handles on the back of the unit .
3 . Check condition of gasket; ensure clamps are out of
the way .
When mounting the eyewash unit to the mounting bracket,
make sure the unit is level and the sprayheads are at a
height of 33-45" from the floor and situated at least 6" from
the wall or nearest obstruction (ANSI Z358.1-2014).
Step 2: Fill and Mount Tank
1 . Pour clean, potable, tepid water into the tank; continue
pouring until the tank is ¹⁄₃ to ½ full .
2 . Add water preservative (not included) to the eyewash
tank . Follow the manufacturer's instructions for the
proper use of water preservative .
3 . Add clean, potable, tepid water to the tank until the water
level reaches the fill line .
4 . Insert one liner into the cap (with hole) and attach the cap
to the top of the tank .
FOR TANK INSTALLATION: To ensure proper installation, the
tank should be placed on the pedestal from the front of the
unit.
5 . Invert the tank and place it on top of the pedestal
as shown on page 5 . The pierce on the pedestal will
puncture a hole in the cap liner and allow solution to
flow .
6 . Raise the arm to snap fit in an upright position . A slight
adjustment to the position of the tank may be required
for proper level placement of the tank on the pedestal .
7 . Secure the tank to the pedestal using the side tank
clamps . Make sure the latches are pulled down tight to
the pedestal (see page 5) .
8 . Lower the arm to activate the water flow . Allow the
water to flow until air bubbles form in the tank . Raise
the arm and "snap" in the upright position . The system
is now ready for use .
After the tank is mounted to the pedestal, minor spillage
may occur. Allow the fluid to stabilize completely before
inspecting the unit for leaks.
4
1/3/2024
33" (838mm)
to 45"
(1143mm)
Cap with Hole
Liner
(insert into cap
before attaching
cap to tank)
Tank
Bradley • 215-1546 Rev. U: ECO 23-05-022B
Installation
44⁵⁄₈"
(1133mm)
to 56⁵⁄₈"
(1438mm)
Turn cap to close