Problems
The sensor does not light up
No return to initial zero
No change in the display or inconsistent
display.
Linearity or accuracy problem.
Copy of the declaration of conformity in the appendix
12. Troubleshooting
Possible causes
Batteries empty.
Fault in circuitry.
Tare function
enabled.
Sensor permanently
deformed due to
incorrect handling,
excessive overload
or compression.
Malfunctioning
sensor or circuitry.
Malfunctioning
sensor or circuitry.
Sensor subject to
compressive or
torsional stress.
13. Declaration of conformity
Remedies
Replace the batteries.
Contact Tractel
Disable the Tare
function and display the
"GROSS" value of the
force.
The product must not be
used before it is verified
by Tractel.
Reset: Switch the sensor
off and then on.
If the malfunction
persists, contact Tractel
Contact Tractel
Eliminate compressive
or torsional stress on the
product.
If the malfunction
persists, contact Tractel
EN
15