Reservoir Float
3. Install the resin cartridge into
the PRIME position to ready for
priming in Step 11.
Step 6: Install I/O Control Wiring
(On-Line and Portable Only)
Refer a qualified electrician to the Analyzer's
Operation and Maintenance Manual for installing
I/O wiring, alarms, and remote start binary input.
Step 7: Install Power
Connect the power cord to Laboratory and Portable
Analyzers.
Or for wiring AC conduit to the On-Line Analyzer, see
the Analyzer's Operation and Maintenance Manual.
Do NOT turn the power on until instructed in Step
11.
Step 8: Install a Printer (Optional)
Install a USB point-of-service (POS) printer, (such as
a Citizen CD-5500A or Epson TM-722OD).
1. Connect the printer cable's USB host connector
into one of the USB host ports on the Analyzer.
2. Connect the cable's USB device connector to the
printer.
3. Configure the printer to the Analyzer later in Step
12.
Step 9: Install the iOS System
(When connecting the Portable or OnLine Analyzer for
online measurements.)
Note: If using a Sievers Autosampler, DO NOT proceed with this
section. Instead, refer to the Sievers Autosampler Installation
Guide.
© 2023 Veolia. All rights reserved.
6
1. Ensure that the power to
2. Verify that the water source
3. Connect the 1/4" PFA inlet
PRIME Position
4. For a two-stream option
5. Slide one end of the waste
6. Ensure that the stainless
7. Measure the flow rate and adjust to between
(When configuring the Laboratory for grab or
Autosampler use.)
1. Using a 7/16" open-end wrench, disconnect the
2. Attach the 1/8" OD plastic waste tubing to the
3. Position the hose for a gravity drain of the waste
the Analyzer is off and
disconnected.
is disabled.
tubing (w/in-line filter) into
the Sample Inlet (A) on the
iOS System. Secure nut with
9/16" open-end wrench. Do
NOT over-tighten.
(M5310 C On-Line Only),
connect the other inlet
tubing into the second
sample inlet.
tubing over the Waste Outlet
Fitting (B) and position the
tubing for a gravity drain.
steel tubing is connected
to the Sample Inlet and
establish a water flow to the iOS System.
50-300 mL/min using the Flow Rate Adjustment
Screw (C).
Step 10: Install the Waste Outlet
nut that connects to the Waste port.
Waste port, then use the 7/16" open-end wrench
to tighten securely.
water.
C
A
B
iOS