8 Technical specifications
8 Technical specifications
28
Programming
Regulation
Output range
Programming range
Output resolution
Programming resolution
Accuracy
Line regulation
Load regulation
Ripple
Short term stability
Long term stability
Start current check
Sudden load change check
Ground leakage check
Output protection
Recovery after power failure
Electrophoresis Power Supply EPS 3501 XL User Manual 18-1130-22 Edition AD
9 programs each with up to 9 phases
Maximum voltage, current and power with automatic
crossover at preset limits
Voltage: 0–3 500 V DC,
Current: 0–400 mA
Power: 0–200 W
Voltage: 35–3 500 V DC
Current: 1–400 mA
Power: 1–200 W
Time: 00:01–500 h
Volthour: 1–500 000 Vh
Amperehour: 1–25 000 mAh
Voltage: 1 V
Current:
1 μA, 0–399 μA
10 μA, 0.4–3.99 mA
100 μA, 4.0–19.9 mA
1 mA, 20–400 mA
Power: 1 W
Voltage: 5 V
Current: 1 mA
Power: 1 W
Time:
1 min, 00:01–99:59,
1 h, 100–500 h
Volthour:
1 Vh, 1–9999 Vh,
100 Vh, 10.0–99.9 kVh
1 kVh, 100–500 kVh
Amperehour: 1 mAh, 1–999 mAh
10 mAh, 1.00–9.99 Ah
100 mAh, 10.0–25.0 Ah
Voltage: 1%, ± 5 V
Current: 1%, ± 1 mA
Power: 2%, ± 1 W
Timer: 0.1% ± 1 min, 00:01–99:59 h
0.1%, ± 10 min, 100–500 h
< 0.2%
< 1% at load change 10–90% of maximum load
< 1% at 3 500 V
< 0.2% /10 h after warm up
< 1% /year
Resistance not greater than 5 M at 40 V (Current less than
8 μA). Can be disabled
Leakage not greater than 500 μA
Fully protected against any overload conditions
Duration < 8 s,
The program continues automatically
Duration ≥ 8 s, The program continues after manual restart