6. Basic Electrical Operations
Refer to Figure A4 and the table below:
Item
Description
Measure current on CH1, Range ± 55 mA (24 V loop
1
on CH2).
Measure current with external loop power (Range:
2
± 55 mA).
3
Measure DC Volts or DC mV on CH1.
4
Measure DC Volts (10 V) or DC mV (10 V) on CH1.
5
Switch test.
7. Pressure Operations
WARNING
Pressurized gases and fluids
are dangerous. Before connecting or
disconnecting pressure equipment, safely
release all the pressure.
CAUTION
To prevent damage to the
pressure station, do not let dirt get into the
pressure mechanism. Before connecting
equipment, make sure it is clean or use the
applicable dirt trap.
7.1 Release the Pressure/Attach the Device
Under Test
1. Open the pressure release valve (one turn) (Ref:
Figure A5).
2. Use the applicable adaptor to attach the device, (Ref:
Figure A7).
7.2 Vacuum Operation
Refer to Figure A6 and the procedure below:
1. Set to vacuum operation (–) (Ref: Item 1).
2. Turn the volume adjuster to mid-range or fully
clockwise (Ref: Item 2).
3. Seal the system (Ref: Item 3).
4. Set the approximate vacuum with the pump (Ref:
Item 4).
5. Adjust the vacuum with the volume adjuster (Ref:
Item 5).
6. On completion of test, open the pressure release valve
one turn (Ref: Figure A5) to release vacuum before
disconnection of device under test.
7.3 Pressure Operation
Refer to Figure A6 and the procedure below:
1. Set to pressure operation (+) (Ref: Item 1).
2. Turn the volume adjuster to mid-range (Ref: Item 2).
3. Seal the system (Ref: Item 3).
4. Set the approximate pressure with the pump (Ref:
Item 4).
5. Adjust the pressure with the volume adjuster (Ref:
Item 5).
6. On completion of test, open the pressure release valve
one turn (Ref: Figure A5) to release pressure before
disconnection of device under test.
Copyright 2014 Baker Hughes Company.
4 | DPI611 Safety and Quick Start Guide–English
8. Specification
8.1 General
Item
Specification
Display
LCD: Color display with touch-screen
Operating temperature
–10 °C to +50 °C (50 °F to 122 °F)
+10 °C to +30 °C (optimized
calibration specification)
0 °C – 40 °C with optional mains
power supply.
Storage temperature
–20 °C to 70 °C (–4 °F to 158 °F)
Ingress protection
IP54 (EN 60529)
Operating humidity
0% to 90% Relative Humidify (RH)
non-condensing.
Shock/Vibration
EN 61010-1, MIL-PRF-28800F
Class 2.
Operating altitude
Maximum 2000 metres (6560 ft)
EMC
EN 61326
Electrical safety
EN 61010-1
Power adaptor
Input range: 100 – 240 VAC, 50 to
60 Hz, 300 – 150 mA, Installation
Category II.
Batteries (Rechargeable) -
Li–ion 14.1Wh Druck Part number:
optional
CC3800GE. Nominal voltage 3.7 V
3800 mAh Charge temperature: 0 °
to 40 °C (32 ° to 104 °F) Discharge
temperature: –10 ° to 60 °C (14 ° to
140 °F). For best battery
performance, keep the temperature
less than 60 °C (140 °F).
Batteries (Non-
8 × AA Alkaline
rechargeable)
Pressure safety
Pressure Equipment Directive -
class: sound engineering practice
(SEP) for group 2 fluids.
Pollution degree
2
Operating Environment
Indoor use only. Not rated for use in
potential explosive atmospheres.
Pressure media
Not rated for oxygen or other
explosive gases or liquids.
8.2 Leak Rates
Model
Leak Rate
DPI611
0.01 bar/min at 7 bar
0.005 bar/min at 95 % vacuum

Publicidad