Important : the data will only be transferred to the application if the application is open while
the device is being used and Bluetooth 4.0 Smart Ready is activated.
The device switches off automatically after eight seconds.
If it is difficult to obtain a reading, try with the other hand or with another finger on the same
hand.
5. CLEANING AND MAINTENANCE
- To clean the silicone inside the oximeter that comes into contact with the finger, use 70° rub-
bing alcohol. Apply the alcohol using a soft clean cloth. Also use the rubbing alcohol to clean
the finger being measured, before and after use.
- Do not pour or spray liquid on to the surface of the oximeter. Make sure that liquid does not
come into contact with the device openings. Leave the oximeter to dry before using again.
- Do not sterilize the device in an autoclave, with ethylene oxide or by immersing it in liquid. The
device must not be sterilized.
6. TECHNICAL SPECIFICATIONS
Product name
BewellConnect® MyOxy
Model
BW-OX1
Classification
Classe IIa
Applied part type BF
IPX22
Continuous use
Display type
LED screen
LED Specifications
Wavelength: Red = 660 ±3nm - Infrared = 905 ±10nm
Radiant power : Red = 3.2mW - Infrared = 2.4mW
Dimensions
32 x 32 x 62 mm
Weight
40 g
Display and
SpO
: Display range : 35 - 99% / Measuring range : 70 - 99%
2
measuring ranges
Pulse: Display and measuring ranges: 30 - 250 bpm
Accuracy
SpO
: 70 - 100% : ± 3%
2
Pulse : ± 2 bpm
Resolution
SpO
: 1%
2
Pulse : 1 bpm
Normal working
Temperature : 5 - 40°C
conditions
Humidity : 15-93%RH
22
BewellConnect - BW-OX1 - User Manual - 062015
Storage conditions
Temperature : -25 - 70°C
Humidity : ≤93%RH
Atmospheric pressure : 86-106Kpa
Technology
Bluetooth 4.0 Smart Ready
Battery
Lithium 280mAh
Autonomy
2 hours in continuous use or 400 separate readings
Automatic shut-down 8 seconds
Bluetooth
10m
transmission distance
Frequency
2400 - 2483.5MHz
USB
USB port 2.0
Accessories provided
Pulse oximeter, instruction manual, quick-start guide, protective pouch,
USB cable
7. TROUBLESHOOTING
If one of the following issues arises when using your MyOxy, consult this trouble-shooting guide.
If the problem continues, please contact our customer service department.
ISSUE
POSSIBLE CAUSE
Sp0
or PRbpm
Your finger must be
2
do not appear
correctly inserted.
correctly
SpO
reading too
2
small to be measured.
Sp0
or PRbpm
Your finger may not
2
are unstable
be inserted deeply
enough.
The patient's finger
is trembling or the
patient is moving.
Impossible to
Battery low
turn on the
Oximeter is damaged
device
BewellConnect - BW-OX1 - User Manual - 062015
SOLUTION
Change the position of your finger and try again.
Try several times. If you have any doubts about a prob-
lem related to the product, seek medical attention
to obtain an accurate diagnosis.
Change the position of your finger and try again.
Try not to move.
Recharge the device with the USB cable
Contact customer support
23