11 Technical Data Continued
Guidance and Manufacturer´s Declaration – electromagnetic Emission – for all
Equipment and Systems
Guidance and manufacturer´s declaration – electromagnetic emission
1
The AC-1000 is intended for use in the electromagnetic environment specified below. The customer
2
or the user of the AC-1000 should assure that it is used in such an environment.
3
Emissions test
RF emissions
4
CISPR11
RF emissions
5
CISPR11
Harmonic emissions
6
IEC 61000-3-2
Voltage fluctuations /
7
flicker emissions
IEC 61000-3-3
Guidance and Manufacturer's Declaration – electromagnetic Immunity – for all
Equipment and Systems
Guidance and manufacturer´s declaration – electromagnetic immunity
The AC-1000 is intended for use in the electromagnetic environment specified below. The customer or the
user of the AC-1000 should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrostatic
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
Compliance
Group 1
Class A
Class A
Complies
IEC 60601
test level
± 8 kV contact
± 15 kV air
± 2 kV for power
supply lines
± 1 kV differential
mode
0 % U
; 0.5 cycle
T
At 0°,45°,90°,135°,
Electromagnetic environment - guidance
The AC-1000 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
The AC-1000 is suitable for use in all establishments, in
cluding domestic establishments and those directly connec
ted to the public low-voltage power supply network that su
pplies buildings used for domestic purposes.
Compliance level
± 8 kV contact
± 15 kV air
± 2 kV for power
supply lines
± 1 kV differential
mode
0 % U
; 0.5 cycle
T
At 0°,45°,90°,135°,
20
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical
commercial
or
hospital