Descargar Imprimir esta página

Alltion AC-1000 Serie Manual Del Usuario página 27

Publicidad

Idiomas disponibles

Idiomas disponibles

Recommended
Communications Equipment and the Equipment or System -for Equipment and
Systems that are not L Life-supporting
portable and mobile RF communications equipment and the AC-1000
The AC-1000 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the AC-1000 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the AC-1000 as recommended below, according to the maximum output power of the
communications equipment
Separation distance according to frequency of transmitterm
Rated maximum
output of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above the recommended separation distance d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Separation
Distances
Recommended separation distances between
150 kHz to 80 MHz
0.12
0.38
1.2
3.8
12
between
Portable
80 MHz to 800 MHz
0.12
0.38
1.2
3.8
12
23
and
Mobile
800 MHz to 2.7 GHz
0.23
0.73
2.3
7.3
23
RF

Publicidad