Dräger-Tubes And Dräger Oil-Impactor - Dräger Aerotest 5000 Instrucciones De Uso

Ocultar thumbs Ver también para Aerotest 5000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39
Use
of the inserted alkaline batteries is insufficient, a warning is displayed
and the alkaline batteries must be exchanged. After that the
measurement is continued.
The Dräger Aerotest 5000 must be switched off while connecting the
Dräger Aerotest 5000 to the mains supply or disconnecting it.
3.2
Dräger-Tubes and Dräger Oil-Impactor
The Dräger Oil-Impactor and the following Dräger-Tubes are used in
combination with the Dräger Aerotest 5000 to measure compressed
gases, especially compressed air. All tube tips must be broken off,
otherwise measurement is impossible. When inserting the tube, the
arrow must point away from the device. Observe the Instructions for
Use provided with the products when using the Dräger-Tubes or the
Dräger Oil-Impactor.
18
Oil-Impactor
Order number
81 03 530/81 03 560
Measuring
0.1/0.5/1 mg/m
range
Colour change
----
Evaluation
After completed measurement, remove the protec-
tive foil from the Impactor and compare the oil con-
centration with the illustration.
3
0,1 mg/m
0,5 mg/m
3
1,0 mg/m
3
00322907.eps
3
Dräger Aerotest 5000

Publicidad

Tabla de contenido
loading

Tabla de contenido