During Use - Dräger Aerotest 5000 Instrucciones De Uso

Ocultar thumbs Ver también para Aerotest 5000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39
3.3.2
Operation/switching off the device
The device is operated via the display (7) using the stylus pin. The
operation is menu-driven. It is used by pressing the display fields or the
arrow to continue and the
menu is accessed by pressing the
is switched off in the main menu by pressing the
NOTICE
i
i
If no menu options are selected in the main menu, the
Dräger Aerotest 5000 automatically switches off after
2 minutes.
3.3.3
Device menus
The following menu options can be selected from the main menu:
Begin Test:
Measurements are performed.
Review Tests:
Measurement results are displayed.
Export Tests:
Measurement results are exported to an
USB stick.
Configure Tester: Device settings are configured.
Via the "Configure Tester" menu option, the language of the display
text can be set to English, German, French or Spanish.
The following functions are available for selection:
set date
set time
set language
set units
update oil list
select standard (appropriate testing standard to test to)
date format
Dräger Aerotest 5000
arrow to go back one step. The main
arrow. The Dräger Aerotest 5000
.
NOTICE
i
i
The settings in the "Set-up Tester" menu shall only be
adjusted by Dräger.
3.4

During use

3.4.1
Measuring
After switching on the Dräger Aerotest 5000, select "Begin Test" from
the main menu. The integrated oxygen sensor is automatically
adjusted to the ambient atmosphere. The device must be located in a
well-ventilated area.
NOTICE
i
i
In temperatures above 40 °C the adjustment takes up to
2 minutes. In high temperature the adjustment may take
longer.
Use the display keyboard to describe the test location. Press the "Fn"
key to enter numbers.
Press the
arrow to continue or the
For most breathing-air standards four test programs can be selected:
Airline < 40 bar,
High Pressure Cylinder ≤ 200 bar,
High Pressure Cylinder > 200 bar,
High Pressure Compressor.
The F3002 pressure reducer with an inlet coupling for an input
pressure of 300 bar max. must be used for all high pressure tests.
Use
arrow to go back.
21

Publicidad

Tabla de contenido
loading

Tabla de contenido