Données techniques
Table 4
type Disjonct. Type Courant Facteur de
I∆N
courant
x0.5
x1
100mA/
x2
300mA/
x5
500mA/
x0.5
1000mA
x1
x2
x5
x0.5
x1
x2
x5
Table 5
type Disjonct. Type Courant Facteur de
I N
courant
x0.5
x1
10mA/
x2
30mA/
x5
100mA/
Palier
300mA/
x0.5
500mA/
x1
1000mA)
x2
x5
Palier
x0.5
x1
x2
x5
Palier
76
Signal de Temps de
Temps de test
prescrit
test I∆N
déclenchement
I∆N
0.5x
-
500ms
1x
130...500ms
500ms
2x
60...200ms
200ms
5x
50...150ms
150ms
0.35x
-
500ms
1.4x
130...500ms
500ms
2.8x
60...200ms
200ms
7x
50...150ms
150ms
0.5x
-
500ms
2x
< 500ms
500ms
4x
< 200ms
200ms
10x
< 150ms
150ms
Signal de Temps de
Temps de test
prescrit
test I∆N
déclenchement
I∆N
0.5x
-
2000ms, I N≥100mA: 500ms
1x
<300ms
500ms
2x
<150ms
150ms
5x
<40ms
40ms
0.4x- 1.4x <300ms
300ms
0.35x
-
2000ms, I N≥100mA: 500ms
1.4x
<300ms
500ms
2.8x
<150ms
150ms
7x
<40ms
40ms
0.1x- 2x
<300ms
300ms
0.5x
-
2000ms, I N≥100mA: 500ms
2x
<300ms
500ms
4x
<150ms
150ms
10x
<40ms
40ms
0.1x- 2x
<300ms
300ms
Norme
A n a l y s e u r
Disjonct.
√
IEC61008-1
√
IEC61008-1
√
IEC61008-1
√
IEC61008-1
√
IEC61008-1
√
IEC61008-1
√
IEC61008-1
√
IEC61008-1
√
0664 P. 100
√
0664 P. 100
√
0664 P. 100
√
0664 P. 100
Norme
A n a l y s e u r
Disjonct.
√
IEC61009-1
√
IEC61009-1
√
IEC61009-1
√
IEC61009-1
√
IEC61009-1
√
IEC61009-1
√
IEC61009-1
√
IEC61009-1
√
IEC61009-1
√
IEC61009-1
√
√
√
√
√