Operation key
MODE Key
Switches the weighing unit.
Shifts to the right the digit flashing when making settings.
ZERO Key
Based on the zero point weighed at the time of turning on the display, by pressing the
ZERO key when the weight value is stable within ±2 % of the weighing capacity, the zero
point indicator turns on as soon as the display becomes zero.
TARE Key
The weight of the item on the weighing pan is subtracted as the weight of a container (tare
weight).
5. Maintenance
・Refer to
"3.
Cautions" regarding use.
・Refer to
"5.3. Error Code
・Refer to the [FG-CWP Series Instruction Manual] (https://aandd.jp/products/weighing/scale/platform/fgcwp.html)
on the A&D website (https://www.aandd.jp) for sensitivity adjustment.
・Periodically check the accuracy of weighing. Adjust the scale, if it is moved to another location or the
environment has changed.
5.1. Repair
Do not disassemble / assemble the scale without an authorized service engineer. Doing so may cause an
electric shock or damage to the scale, etc. In this case, repair is not covered under warranty. Contact your
local A&D dealer if your scale needs service or repair.
5.2. Error display
In this situation
Nothing displayed.
Scale does not turn on.
The scale does not display
zero at first.
is displayed.
is displayed.
is displayed.
is displayed.
is
displayed and does not
proceed.
Table" and corresponding mode for displayed error code.
・Is the main power the correct voltage?
・Is the main power cord properly connected?
・Check around the weighing pan.
・Is there anything on the weighing pan?
・Refer to the [FG-D-CWP / FG-D-ACWP Instruction Manual]
(https://aandd.jp/products/weighing/indicator/fgdcwp.html) on the A&D
website (https://www.aandd.jp) and perform zero point sensitivity
adjustment.
・Weighing error that meaning "Over loaded."
・Indicates that the sample weight is too light to set the unit weight in the
counting mode.
・Sensitivity adjustment error that meaning "Too light."
・Displayed when the main power drops.
・The weighed value is unstable due to drift, vibration or other factors.
A breeze or vibration may be affecting measurement.
Check around the weighing pan.
・Check the connection of the load cell cable.
Description
Confirm these items