Trouble shooting
Error Message
ENGLISH
Problem
The probe cover has not been
installed well. will flash
LCD screen and can't take the ear
temperature
Error 5~9, the system is not
functioning properly.
Measurement before device
stabilization.
The device showing a rapid
ambient temperature change.
The ambient temperature is not
within the range between 10 °C
and 40 °C (50 °F ~104 °F).
Temperature taken is higher than
+42.2 °C (108°F)
Temperature taken is lower than
+34 °C (93.2 °F)
Device cannot be powered on to
the ready stage.
Solution
Please check the setting of the
probe cover again.
on the
(Refer to the section of "Place
the Probe Cover" )
Unload the battery, wait for 1
minute and repower it. If the
message reappears, contact the
retailer for service.
Wait until all the icons stops
flashing.
Allow the thermometer to rest in
a room for at least 30 minutes at
room temperature: 10 °C and 40
°C (50 °F ~104 °F).
Check the integrity of the
probe cover and take a new
temperature measurement.
Make sure the probe cover
i s c l e a n a n d t a k e a n e w
temperature measurement.
Change with a new battery.
12