Control Testing
Applying the control solution
8 Apply a large hanging drop of control
solution directly from the bottle to the test
strip. Ensure that neither the bottle nor the
finger touches the application area. The
application area must be completely
covered.
You can also apply the control solution
outside the meter. The procedure is the
same as when you apply a blood sample.
Refer to the description starting on
page 48.