Product description
Note
Only use the IEM NiMH power packs or alkaline batteries provided. Although zinc-carbon batteries and NiCd
power packs show sufficient voltage during the battery test, the power is often insufficient to take
measurements for 24 hours. The power packs should be discharged and recharged several times before their
first use. Please observe the enclosed operating instructions for the charger.

Switching on

Always check the condition of your blood pressure monitor first before giving it to a patient. This is done by
observing the first notifications on the display of the blood pressure monitor shortly after switching it on.
The following should be displayed in this sequence:
Test
Battery status (volts)
Display Segment Test
Current 24-hour period
If an error occurs during the internal test, the blood pressure monitor will show E004 on the display and an audible
signal will sound. For safety reasons, the blood pressure monitor will be out of operation. Send the blood pressure
monitor immediately to your specialist supplier or directly to IEM GmbH for repair.
86
Display
Comment
2.85
at least 2.6 V for NiMH power packs
at least 3.10 V for alkaline batteries
999:999
First the digits (999:999 to 000:000), then all other symbols of the
to
LCD appear in succession.
000:000
Check whether all segments are correctly and fully displayed (the
complete program code is checked for correctness in the
background).
21:45
Between 00:00 and 23:59