natus Camino 110-4HMT Instrucciones De Uso página 8

Equipo para la monitorización de la presión y la temperatura microventricular con perno
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 27
INTERNATIONAL MARKINGS
CATHETER EXTENSION CABLE
CONNECTIONS
Symbol also signifies:
TYPE BF EQUIPMENT: Protected against
electric shock, having an F-type isolated
(floating) applied part.
Sterile - Sterilized using
Ethylene Oxide.
Use by.
Do not reuse.
Manufacturer.
2
Do not resterilize.
STERILIZE
MR Unsafe.
MR
REF
Catalog Number.
Ventricular Catheter:
Outside Diameter
Inside Diameter
Length
Transducer size
Transducer type
Frequency response (system)
Model MPM-1/SPM-1/CAM01/CAM02
Model V420/M420/420XP
Model 420
Measurement range (system)
Zero drift (system)
First 24 hours (maximum)
5 days (typical)
Reference pressure
Overpressure
Temperature coefficient
Linearity and hysteresis (system)
Pressure range:
-10 to 50 mmHg
51 to 125 mmHg
Temperature Measurement
Temperature Measurement Resolution
Temperature Measurement Range
Temperature Measurement Accuracy
MANUFACTURING INFORMATION
Integra NeuroSciences
5955 Pacific Center Blvd
San Diego, CA 92121 - USA
6
(floating) applied part; intended for direct
cardiac application.
sale by or on the order of a physician.
SPECIFICATIONS
3.7 mm
2.2 mm
6-8 cm (adjustable)
4F (1.35mm Diameter)
Fiber optic
100 Hz (-3 dB)
120 Hz (-3 dB)
33 Hz (-3 dB)
-10 to 125 mmHg
0 ± 2 mmHg
less than ± 1 mmHg per day
Atmosphere
-700 to 1250 mmHg
Max of 3 mmHg over temperature range of
22°C to 38°C (70°F–100°F)
± 2 mmHg or better
± 6% of reading or better
0.1°C
30°C–40°C
± 0.3°C (30°C–40°C)
EC REP
Type CF Equipment: Provided with
a degree of protection against electric
shock higher than Type BF isolated
Caution.
Lot Number.
Caution: Federal (USA) law
restricts this device to
Authorized representative in
the European Community.
Contains or presence of DEHP.
Consult Directions for Use.
Integra LifeSciences Services France
Immeuble Séquoia 2
97 Allée Alexandre Borodine
Parc Technologique de la Porte des Alpes
69800 Saint Priest, France
Tel: 33 (0) 4 37 47 59 10 Fax: 33 (0) 37 47 59 29