Hach si792 D Manual De Usuario página 63

Ocultar thumbs Ver también para si792 D:
Tabla de contenido

Publicidad

Tabla 19 Parámetros del bus del bloque de transductores (continuación)
Parameter
TRANSDUCER_
DIRECTORY
TRANSDUCER_TYPE
XD_ERROR
COLLECTION_
DIRECTORY
Output
INPUT_RANGE
POLARIZATION_VOLTAGE Sets the polarization voltage
SENSOR_TYPE
Description
Directory that specifies the
number and the starting indices
of the transducers in the
transducer block.
Identifies the transducer type.
A transducer block sub-code.
XD_ERROR contains the
highest priority alarm that has
been activated in the
TB_DETAILED_STATUS
parameter.
A directory that specifies the
number, starting indices, and
DD item of IDs of the data
collection in each transducer
within a transducer block. Used
by the host for efficient transfer
of information.
Selects the connection of the
sensor
Selects the used sensor type.
Data
R/W Bytes
Range
type
R
4
Default: 65535 =
R
2
other
R
1
Default: 0
R
36
0 = LO_LEVEL
R/W 2
uns16
1 = HI_LEVEL
R/W 4
float
–1000–0
0 = Standard
R/W 2
uns16
1 = Traces

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Si792x dSi792x d-ffSi792x d-pa

Tabla de contenido