Mettler Toledo BTA 221 Manual De Usuario página 19

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

08/10
Order number 72244520
F1.3.3 – Adjustment
The steps with grey background only appear if parameter
F1.3.2=LinOn is set.
Display
Key
E SCL
10 CAL
...
0 CAL
Add Ld
000000
003000
10 CAL
...
0 CAL
FULL Ld
000000
006000
10 CAL
...
0 CAL
donE
F1.4
Description
Unload pallet truck scale
Confirm empty pallet truck scale
The weighing terminal counts down from 10 to 0.
Zero is determined
Load half of the maximum load
Confirm half load
Enter weight value for half maximum load
Enter weight value
Weight value for half of the maximum load entered
Confirm weight value
The weighing terminal counts down from 10 to 0.
Half maximum load is adjusted
Load maximum load
Confirm maximum load
Enter weight value of maximum load
Enter weight value
Weight value for maximum load entered
Confirm weight value
The weighing terminal counts down from 10 to 0.
Maximum load is adjusted
Adjustment finished. This message is displayed
for about 2 seconds
Next block in the supervisor menu
19
User Manual BTA221
METTLER TOLEDO

Publicidad

Tabla de contenido
loading

Tabla de contenido