Sartorius Entris II Advanced Line BCA Instrucciones De Manejo página 44

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 161
Malfunctions
Warning message
Fault
ERR 10
Taring is not possible .
ERR 11
The weight value
cannot be saved in
the tare memory .
Low voltage of
battery of the
internal clock
module: The date
and time may be
incorrect .
Calibration
The device cannot be
canceled
adjusted when
starting the
calibration function
due to a zero point
error or an absence of
leveling (conformity-
assessed models) .
The balance
needs to be
adjusted!
Remaining time in
minutes: xx
Weight value is
invalid
The balance
needs to be
leveled!
44
Entris
II Advanced Line Operating Instructions
®
Cause
The device cannot be
manually tared because
an application program
has the tare memory
reserved .
The weight value is
negative or "zero" .
The battery is nearly
empty .
The device was not
zeroed before calibration .
The device is loaded .
The balance drifts too far .
The internal motor weight
unit is defective .
Conformity-assessed
models: The device is not
leveled .
The device needs to be
calibrated and adjusted .
The device is in the warm-
up phase and has not yet
reached its operating
temperature .
xx = remaining time in
minutes
The displayed value is
invalid .
The device is not leveled .
Remedy
To release the tare memory:
Exit the application
program with the [End]
button .
Check the sample being
weighed .
Zero the device before
placing the sample on the
balance .
Contact Sartorius Service .
Zero the device .
Check the preload and set if
necessary .
Remove the sample from
the weighing pan .
Allow to warm up . Re-align
the device .
Contact Sartorius Service .
Calibrate and adjust the
device .
Comply with the warm-up
time after switching the
device on .
Zero the device .
Level the device .
Chapter,
page
7 .2, 37
14 .8, 49
8 .4, 39

Publicidad

Tabla de contenido
loading