12. PERIODIC SAFETY CHECKS
We recommend that you inspect the device
for mechanical and functional damage and the
safety labels for legibility annually/every 12
months.
Applicable in Germany only: Messtechnische
Kontrolle nach MPG (Medizinproduktegesetz)
alle 24 Monate.
13. MAINTENANCE
Only the manufacturer or a certified service
center can carry out any service and repairs
other than those mentioned above. Before
sending the device for servicing, make sure that
you have stored your measurement data in a PC
onto which LINK software has been installed.
14. CHARGING THE BATTERY
When the battery is low, an error message will
indicate that you must recharge the battery.
A full charge takes approximately one hour. If
a green light is blinking on the upper navigation
button, this means that the tonometer is
charging. When the green light is constant,
the device is completely charged.
15. ACCESSORIES
Part number Product Description
542
Probe base + probe holder
7215
Probe base collar
572
Docking station
573
Micro USB charger
575
USB cable
520
Aluminum case
543
Probe base cleaning container
120
Probe box
16. TECHNICAL DATA
• Type designation TA03
• The device conforms to CE regulations
• Dimensions: 225 mm x 46 mm x 90 mm (8.9'' x 1.8'' x 3.5'')
• Weight: 275 g (9.7 oz.)
• Power supply: Rechargeable internal polymer Li-ion battery 3.7V 480mAh. Up to 500 measurements
can be made with a fully charged battery.
• Measurement range: 5-50 mmHg
8
There are four alternative ways of charging the
battery:
1. Connect the USB cable between the Icare
PRO and the charger. Select a suitable plug
from the alternatives and attach it to the
charger. Connect the charger to the mains.
When charging is complete, disconnect the
charger from the mains (Figure 10).
2. If you have the optional docking station, use
the USB cable to connect the USB charger to
the docking station. Insert the Icare PRO into
the docking station. Connect the charger to
the mains. When the Icare PRO is completely
charged, disconnect the charger from the
mains (Figure 11).
3. Turn on a PC that has the Icare LINK software
installed and running. Keep the PC and Icare
PRO 1.5 m or more away from the patient.
Use the USB cable to connect the Icare PRO
to the PC, in order to charge the Icare PRO
using the PC (Figure 12).
4. Turn on a PC that has the Icare LINK software
installed and running. Use the USB cable to
connect the optional docking station to the
PC. Keep the PC and Icare PRO 1.5 m or more
away from the patient. Insert the Icare PRO
into the docking station to begin charging it
using the PC (Figure 13).
Weight
Dimensions
4g
7 mm x 34 mm
1g
14 mm x 12 mm
900g
190 mm x 120 mm x 50 mm
118g
Charger Unit: 58 mm x 37 mm x 44 mm
US/Japan: 30 mm x 40 mm x 26 mm
EU: 37 mm x 40 mm x 44 mm
Australia: 40 mm x 33 mm
UK: 47mm x 50 mm x 36 mm
23g
1 m
1050g
335 mm x 265 mm x 90 mm
3g
5,6 cm x 2 cm
9g
10,1 cm x 4,6 cm x 1,5 cm